Alternative names, keywords |
Co, metal on metal, MoM |
Samples required |
Collect two samples and send the second for cobalt and chromium testing. The first drawn sample is used to wash out any potentially contaminating traces of metals from the steel needle and can be used for other assays (e.g. FBC). Chromium and cobalt can both be done on the same sample, but due to the risk of metal contamination the sample cannot be shared with any other tests. if querying deficiency or toxicity from other cause than MoM hip replacement, please send a second-drawn clotted blood for serum cobalt:
|
Test indications |
Investigation of metal-on-metal hip replacements. |
Reference range |
In MoM hip replacement, MDA/2017/018 action limit post-implant is >119 nmol/L (7 ppb). See MDA/2017/018 (issued 29 June 2017) for guidance on chromium and cobalt testing in MoM hip replacements and management. To convert from nmol/L to µg/L (ppb): nmol/L x 0.059 = µg/L |
Turnaround time |
Referral lab quotes turnaround time of 5 days, however additional time must be allowed for packing, dispatch and delivery of sample, and return and processing of results. |
Enquiries |
Biochemistry (Referrals) |