Alternative names, keywords |
|
Samples required |
24 hour urine collection in plain container. Instructions for 24h urine collection available here or on UHSussex public site. See also protein: creatinine ratio. |
Test indications |
|
Factors affecting test performance/ results |
Haemoglobin interferes. Levodopa, methyldopa and Na2-cefoxitin cause artificially high total protein results at the therapeutic drug level. Phenazopyridine and calcium dobesilate cause artificially low protein results at the therapeutic drug level. Patient samples containing > 8 g/L of organically bound iodine from radiopaque media (e.g. Hexabrix) may have falsely elevated results. In alkaptonuria, homogentisic acid in urine samples at concentrations > 0.6 mmol/L can cause false results. The administration of gelatin-based plasma replacements can lead to increased urine protein values. Results should always be assessed in conjunction with the patient’s medical history, clinical examination and other findings. |
Reference range |
< 140 mg/24h |
Turnaround time |
Same day |
Enquiries |
Biochemistry (Automation) |